Q Why I am getting split peaks on my chromatograms?

A Split Peaks

 

What you expect What you got

 

Possible Cause

Remedy

Contamination of column or guard

Remove guard column. If the problem is solved replace it. If not the problem is most likely due to a blocked column frit which may need replacing. Otherwise replace the analytical column.

Sample Solvent incompatible with mobile phase

Change solvent used in sample preparation. This will ideally be the mobile phase.

Q Why I do I have a noisy baseline on my chromatograms?

A Base line noise (Irregular)

What you expect What you got

Possible Cause

Remedy

Leak

Check for any evidence of leaking and fix as appropriate.

Poor quality mobile phase

Use HPLC grade solvents

Cell contaminant

Clean detector cell

Weak lamp

Replace detector lamp

Air Bubbles in detector

Purge Detector

 

 

 