Sitemap: Home

ngimg0






Lost Password?
No account yet? Register
 
 
 

 
Chroma Forum
Chroma Forum  
Home | post reply | threaded view | help
::post new topic::
Author Message
Agilent 1100 column heater - 2005/06/17 07:04 The 1100 series column heater heats the mobile phase and not the air surrounding the column. This could in theory affect the tranfer of methods from or to 1100 series instruments. Jost column ovens that I have come across heat the air rather than the mobile phase. Another temperature consideration for the 1100 series is which side of the heater block you run the solvent through. There is a 3uL side and a 6uL side. I typically run through the 6uL side for better heating of the mobile phase

Post edited by: admin, at: 2005/06/17 07:04
  reply | quote
::post new topic::

© 2008 LC Services
HPLC l Service Contracts l Gas Chromatography l Mass Spectrometry l Qualification l Breakdowns l Supported Models l Company Overview
Chorma Forum l Parts and Accessories l Refurbished Equipment l HPLC Training l Sitemap l Contact Us

Registered office: First floor, 5 Doolittle Yard, Froghall Road. Ampthill, Beds. MK45 2NW
Registered in England & Wales No.3219349 - VAT Number GB705847031
BS EN ISO 9001: 2000 Registered Cert No. 18198 / UKAS Quality Management