LC Services
Sitemap: Home

ngimg0






Lost Password?
No account yet? Register
 
 
 

 
Chroma Forum
Chroma Forum  

Post a new message in"HPLC Troubleshooting"
name:
Subject:
topic icon:
no
boardcode:
 
 Colour:  Size: Close all tags
Message:

emoticons:
B) ;) :) :laugh:
:ohmy: :x :mad: :blink:
:P :( :unsure: :rolleyes:
:y32b4: :lol: :dry: :huh:


boardcode is enabled
HTML is disabled
* - your email address will never be displayed on the site.

Topic History of: Loss of resolution through heating.
Max. showing the last 6 posts - (Last post first)
Author Message
admin I have encountered different opinions
concerning buffered mobile phase pH. One side
says to make the buffer solution to the
required strength and pH and then add to the
organic. The other side says to make the
buffer solution to the required strength and
add to the organic; then, you adjust the pH.
Which is the proper procedure? It would
appear to me that you would have to adjust
the buffer first. Thanks for any input.
admin When I was writing about the use of a
circulating water bath to control
temperature, I forgot to note that in this
case around 25 cm of metal capillary with
small ID should be in the waters bath too.
BTW: Merck recommended for the use of
Chromolith columns to use a capillary with
0.2 mm ID and lenght of at least 30 cm for
flow rates below 3.0 ml/min and at least 40
cm for low rates above 3.0 ml/min.

Kind regarts, Andreas
admin Some time ago I run into such heating
problems. My customer used an HPLC system
with different setups. Column temperature was
50 degr. C! He used for QC an analytical
column, a semiprep and a prep column,
\"same\" method, same compounds. All columns
where packed with the same material, of
course the prep had a bigger particle size.
Peak shape was excellent on both analytical
and semiprep column, but the prep column
showed dubble peaks, peak sholders and a bad
resolution. So we exchanged the prep column.
But the second prep column showed the same
thing. Customer was upset, and I visited his
lab, to connect the second exchange column.
Autosampler was checked, pumps and detector
aswell. Everything was ok. But also the third
prep column showed bad peak shape, peak
sholders etc.! At the end we found out,
capillary between autosampler and column, in
the oven, was too short for the prep column.
Running with high flow rate eluent had no
time to have the same temperature than the
column. So we had a temperature gradient on
the column. Solution was to pre-heat eluent
and to use a longer capillary. Now also the
prep column showed an excellent performance,
peak shape was ok, resolution was ok and
customer was happy again.
admin Heating the mobile phase could result under
some circumstances in the loss of resolution
(compared to the unheated mobile phase):
Friction leads to higher temperature in the
inner area of the column than in the outer
area and to an acceleration of mobile phase
(and parts of the deluted sample).
Let me simplify the mechanism: We have an
outer (stationary) phase which is cooler
(cooled from outside) than the inner
(stationary) phase (heated up by friction)
due to the different proximities to the wall.
As the solvent passes through the column the
outer mobile phase slows down while the inner
mobile phase continues at fast speed. The
outer mobile phase passes through in a
round-about way while the inner mobile phase
passes directly. At the entrance friction
remains constantly high. The higher the
pressure the higher the friction. The higher
the friction the higher the heat. The higher
the heat the lower the viscosity. The lower
the viscosity the faster the speed.
This results in peak broadening.
Of course, this happens also at ambient
temperature and rises with increased pressure
and flow rates.
If the mobile phase has a lower temperatur
than the column the higher viscosity slows
down the inner mobile phase and peak
broadening will be reduced.

Also in some cases it is not possible to
transfer temperature sensitive methods from a
hplc system with heating blocks (like the
agilent systems) to systems with air heating,
because the selected and the resulting real
temperature of one hplc system vary too much
and interlab reproducibility (with different
hplc) is even worse (as much as 10°C
difference!).
Best way to reproduce heating a column is to
use a circulating water bath and extra
temperature control (± 0.1°C). Normally this
is not realistic in practice.
It should be good lab practice - however not
a common one - to report the data of the
thermostat used in an hplc system.
Always give the column enough time to adapt
to non-ambient temperature. The wider the
column id the longer the waiting time.

On the other hand, most methods I've seen
produce the same results at ambient
temperature as at 30°C.
And even if the built-in thermostat of a hplc
system shows some diviation between selected
und resulting real temperature these effects
would not be as high as the ambient
temperature changes inside some labs
(especially if in the same room one or more
gc systems are running as well).
So in most cases using a thermostat is not a
mistake.

Peak broadening could be reduced by
increasing the column temperature (and so
reducing back pressure and the viscosity of
the mobile phase).
Also, it could be reduced by changing the
dimensions of the column; for example instead
of using the typical column (250 x 4.6 mm) I
suggest unsing a shorter column with wider id
(150 x 6 mm). Or you can reduce the lenght
and flow rate by half if seperation allows
this.
Even when using a column with 150 x 6 mm and
3 µm particles instead of 250 x 4.6 mm and 5
µm particles, back pressure will decrease and
often results in smaller and higher peaks.
However, be carefully, higher peaks could
disturb the linearity.

Bye, Andreas Neumaier

© 2009 LC Services
HPLC l Service Contracts l Gas Chromatography l Mass Spectrometry l Qualification l Breakdowns l Supported Models l Company Overview
Chorma Forum l Parts and Accessories l Refurbished Equipment l HPLC Training l Sitemap l Contact Us

Registered office: First floor, 5 Doolittle Yard, Froghall Road. Ampthill, Beds. MK45 2NW
Registered in England & Wales No.3219349 - VAT Number GB705847031
BS EN ISO 9001: 2000 Registered Cert No. 18198 / UKAS Quality Management