Topic History of: Hp g1800a gcms
Max. showing the last 6 posts - (Last post first)
Author Message
admin Hello. Would like to ask for help regarding
this problem.

Whenever I tried to autotune for my GCMSD, I
will obtained a high EMV of 2300+. This
arises whenever I used a higher column flow
(in this case, 2mL/min). When i tried to
lower the column flow to..ie 1.8mL/min, I
will obtained a lower EMV. The question is,
is there a problem with my column?Is it
deteriorating or is there any other
problem???What can be done to correct this
problem?

FYI, I have cleaned the ion source but I knew
that this is not the problem due to the fact
that I obtained lower EMV with lower column
flow.

Thank yopu for your help and kind assistance.
admin Looking for information on g1800a GCD
In particular the service level password
software is original windows 3.1

© 2009 LC Services
HPLC l Service Contracts l Gas Chromatography l Mass Spectrometry l Qualification l Breakdowns l Supported Models l Company Overview
Chorma Forum l Parts and Accessories l Refurbished Equipment l HPLC Training l Sitemap l Contact Us

Registered office: First floor, 5 Doolittle Yard, Froghall Road. Ampthill, Beds. MK45 2NW
Registered in England & Wales No.3219349 - VAT Number GB705847031
BS EN ISO 9001: 2000 Registered Cert No. 18198 / UKAS Quality Management