Agilent 1100 column heater - 2005/06/17 07:04 The 1100 series column heater heats the mobile phase and not the air surrounding the column. This could in theory affect the tranfer of methods from or to 1100 series instruments. Jost column ovens that I have come across heat the air rather than the mobile phase. Another temperature consideration for the 1100 series is which side of the heater block you run the solvent through. There is a 3uL side and a 6uL side. I typically run through the 6uL side for better heating of the mobile phase

Post edited by: admin, at: 2005/06/17 07:04